Screen size :
3.5",5.6”,10.4”
Screen type: Color TFT LCD with resistive analogue
Touch-Panel, or blue screen or black screen
Screen resolution : (620 x 480 pixels) or better
Number of Input channels : 12 or more
Inputs data
-
1. Type
of inputs: dc volts/dc mv/dc mA/TC J, K, R, S & N type),2/3 Wire RTD(at least
PT100, PT1000)
-
2. Mix
of inputs : freely
configurable
-
3. ADC
resolution : 16 bits or better
-
4.Input
update rate : 125 ms or better on all
channels
-
5. Trend
update rate : 125 ms or better
-
6.CMRR : 100dB or better
between channel to (48-53 Hz or wider) channel
or channel to ground
-
7.Common
mode voltage : 230 Volts or more
(continuous)
-
8. Isolation :
-
Channel
to channel : 300V or more RMS and dc
-
Channel
to ground : 300V or more RMS and dc
-
9. Input
ranges and accuracies (worst case at 20 Deg C):
-
i.Volts
(-35
to +35) +/-5 mV(0.03% I/P + 0.06% or better)
:(-150
to +150) +/-10 mV(0.03% I/P +
0.04% or
better)
:
-1 to +1 V (0.03% I/P + 0.03% or better)
:-20
to +20 volt (0.4% I/P + 0.035% or better)
Up to
1 volt range : 25 ppm or better
For
20 volt range : 400 ppm or better
-
ii.Accuracy degradation with temperature
(for volt input)
-
iii.Thermocouples
-
Cold
junction types : Off, internal, external,
remote
-
CJ
error: 1 deg C max (inst. at
25 degC)
-
Range
and max linearization errors:
-
a.Type
J : 20 to 1200 degC(0.02degC or better)
-
b.Type
K: 20 to 1372 degC(0.04degC or better)
-
c.Type
R: 20 to 1768 degC(0.04degC or better)
-
d.Type
S: 20 to 1768 degC(0.04degC or better)
-
e.Type
N: 0 to 1300 degC(0.04degC or better)
-
iv.RTD range and max linearization error:
-
a.Range : 20 to 850 Deg C (0.01 degC or better)
Paperless Recorder Interfaces:
-
1. USB : with support for at least
16 GB pen drive
-
2. CF
card slot: with support for
at least 16 GB CF card)
-
3. Ethernet
slot
-
i. Type:
10/100baseT Ethernet (IEEE802.3)
-
ii. Protocols: at least TCP/IP, FTP, EtherNet/IP server
-
iii. Cable Type : CAT5
-
iv. Cable length:
100 meters or more
-
v. Termination : RJ45
Functions
-
Events
-
i. Nos. :
at least 3 configurable
-
ii. Sources for events : Batch start, Batch running, Batch
-
Stopped, Alarm active, Archive
-
Media threshold crossed
-
iii.Logical operations on events: AND, OR
-
Paperless Graphic recorder MATH functions
-
i. Virtual channels
-
ii. Functions required: add, multiply,
divide, subtract, constants, Log, e to the power,
-
10 to the power, rate of change,
-
Group average, group minimum, Group maximum
-
iii. Utility : all utilities with physical channel
-
i.e. display, Trending, logging, bar chart are
available for virtual channels too.
-
Custom linearization for customized
linearization of inputs
-
i. No. of tables : at least 6
-
ii. No.
of entries per table : at least 32
-
Displays
and features:
-
i. Bar
graphs with individual scales and value display
-
ii. Current
trending with cursor to know value at any position
-
iii. Historical
trending with cursor to know value at any position
-
iv. Both
hoz and vertical trending possible.
-
Real
Time clock
-
i. It
is available
-
ii. With
battery backup for at least 6 months with recorder unpowered.
-
Alarms
for low and high with alarm logging is available for each channel.
Software
(in scope of supply)
-
1.To
be installed on a PC for analyzing the logged data. PC is not in scope of
Supply.
-
2.It
is possible to network at least 5 paperless recorders and connect to software.
-
3. It
is possible to view standard screens of all networked recorders dynamically.
-
4.It
is possible to view recorded data & reports.
Relay
Outputs
-
1.Numbers : at least 3 C/O type
-
2. Update
rate : at least 500 ms
-
3. Contact
current rating : 1 Amp or more
-
4. Contact
voltage rating : 250 V or more
-
5. Switching
VA: 200 VA or more
-
Safety isolation: 300Vrms or dc between relay to relay
Data
storage/Archive
-
1. Internal
flash memory
-
2. On
removable media
-
3. Over
Ethernet
-
:96
MB or more
-
:CF
card/USB drive configurable for on demand or periodic transfer from internal to
external memory.
-
:Configurable
for archiving in the connected computer
Environment:
-
1. Temperature
Operation : 20 to 50 degC or
wider
-
2. Humidity
limits Operation : 5% to
80% RH
-
3. Protection
Bezel and display: IP65
-
4. Vibration
(10 to 150Hz) : BSEN60873 Or IEC60068-2-6