The flow range applies to Silver Automation Instruments SRK-100 series thermal mass flow meters.
1. Flow range Table Reference
Table 1: Air mass flow meter flow range selection table
Table 2:Air, N2 (nitrogen) ,O2 (oxygen gas),H2 (hydrogen) Flow Meter Flow range table
Table 3: Argon (Ar)gas, Helium(He)gas, Natural Gas Methane ,LPG, City Gas, Chlorine Gas flow measurement flow range table
All thermal dispersion mass flow meters are calibrated before leaving factory, the upper limit of flow range will be indicated on the name plate of flow meter.
2. Lower and upper limit of the flow range
1) Lower limit: the thermal mass flow meter is sensitive to low flow, it can detect 0.05 m/s flow rate, so in normal condition, lower limit can be ignored.
2) Upper limit: It is normally decided by process design; the flow meter upper limit is more than 20% of design limit.
3) Mixed gas measurement: For mixed gas, the ender user should provide standard density and mole ratio (different gas percentage of the mix gas). The measuring range is decided by our factory. It is normally calibrated by air or nitrogen flow, and then corrected by meter conversion coefficient.
3. Statement
The flow range is limited by equipment; please get confirmation from factory before purchasing.
Air mass flow rangeselection table
DN (mm) |
Min (kg/h) |
Normal(kg/h) |
Max(kg/h) |
25 |
0-1.14 |
0-78 |
0-204 |
40 |
0-2.92 |
0-194 |
0-525 |
50 |
0-4.57 |
0-310 |
0-822 |
80 |
0-11.7 |
0-775 |
0-2106 |
100 |
0-18.3 |
0-1292 |
0-3291 |
150 |
0-41.1 |
0-2584 |
0-7404 |
200 |
0-73.1 |
0-5168 |
0-13613 |
250 |
0-114 |
0-7752 |
0-20564 |
300 |
0-165 |
0-11162 |
0-29613 |
400 |
0-292 |
0-19845 |
0-52645 |
500 |
0-457 |
0-31008 |
0-82257 |
600 |
0-658 |
0-44652 |
0-118450 |
700 |
0-896 |
0-60775 |
0-161224 |
800 |
0-1170 |
0-79380 |
0-210578 |
900 |
0-1481 |
0-100465 |
0-266513 |
1000 |
0-1828 |
0-124032 |
0-329028 |
1200 |
0-2632 |
0-178606 |
0-473801 |
1500 |
0-4113 |
0-279072 |
0-740314 |
2000 |
0-7312 |
0-496128 |
0-1316113 |
Four kinds common gas calibrated range selection table
DN(mm) |
Air |
(N2) |
(O2) |
(H2) |
25 |
60 |
60 |
60 |
28 |
40 |
150 |
150 |
150 |
70 |
50 |
240 |
240 |
240 |
112 |
80 |
600 |
600 |
600 |
282 |
100 |
1000 |
1000 |
1000 |
470 |
150 |
2000 |
2000 |
2000 |
940 |
200 |
4000 |
4000 |
4000 |
1880 |
250 |
6000 |
6000 |
6000 |
2820 |
300 |
8640 |
8640 |
8640 |
4060 |
400 |
15360 |
15360 |
15360 |
7219 |
500 |
24000 |
24000 |
24000 |
11280 |
600 |
34560 |
34560 |
34560 |
16243 |
700 |
47040 |
47040 |
47040 |
22108 |
800 |
61440 |
61440 |
61440 |
28876 |
900 |
77760 |
77760 |
77760 |
77807 |
1000 |
96000 |
96000 |
96000 |
45120 |
1200 |
138240 |
138240 |
138240 |
64972 |
1500 |
216000 |
216000 |
216000 |
101520 |
2000 |
384000 |
384000 |
384000 |
180480 |
Note: The flow unit in above table is Nm3/h, corresponding flow rate is 34m/s,when in application, it can expand to 50m/s
Common gas upperlimit (Nm3/h)
DN (mm) |
Argon (Ar) |
Helium (He) |
Natural Gas |
Methane Gas |
LPG |
City Gas |
Chlorine Gas |
25 |
82 |
37 |
58 |
51 |
84 |
44 |
109 |
40 |
207 |
94 |
147 |
129 |
211 |
111 |
273 |
50 |
331 |
151 |
235 |
206 |
339 |
177 |
436 |
80 |
828 |
378 |
588 |
516 |
847 |
444 |
1092 |
100 |
1380 |
630 |
980 |
860 |
1143 |
740 |
1820 |
150 |
2760 |
1260 |
1960 |
1720 |
2826 |
1480 |
3640 |
200 |
5520 |
2520 |
3920 |
3440 |
5652 |
2960 |
7280 |
250 |
8280 |
3780 |
5880 |
5160 |
8478 |
4440 |
10920 |
300 |
11923 |
5443 |
8467 |
7430 |
12208 |
6393 |
15724 |
400 |
21196 |
9676 |
15052 |
13209 |
21703 |
11366 |
27955 |
500 |
33120 |
15120 |
23520 |
20640 |
33912 |
17760 |
43680 |
600 |
47692 |
21772 |
33868 |
29721 |
48833 |
25574 |
62899 |
700 |
64915 |
29635 |
46099 |
40454 |
66467 |
34809 |
85612 |
800 |
84787 |
38707 |
60211 |
52838 |
86814 |
45465 |
111820 |
900 |
107308 |
48988 |
76204 |
66873 |
109874 |
57542 |
141523 |
1000 |
132480 |
60480 |
94080 |
82560 |
135648 |
71040 |
174720 |
1200 |
190771 |
87091 |
135475 |
118886 |
195333 |
102297 |
251596 |
1500 |
298080 |
136080 |
211680 |
185760 |
305208 |
159840 |
393120 |
2000 |
529920 |
241920 |
376320 |
330240 |
542592 |
284160 |
698880 |
Standard state: temperature 0 ℃, Pressure 1.01325×105Pa (absolute pressure)
Flow unit option: kg/h,t/h,Nm3/h,Nkm3/h.
1. Recommendation:
When the pipeline≤Ø100, Flange thermal mass flow meter is recommended.
When the pipeline >Ø100,Insertion thermal mass flow meter is recommended.
2. If the field pipes have been installed, no installation flanges available, Insertion type can be selected. Please specify when ordering.